Publication Date: 2023/11/29
Abstract: The influence of flow rate and voltage in the Double Dielectic Barrier Discharge (DDBD) reactor in producing medical ozone has been carried out. The aim of this study was to obtain medical ozone as a standard dose for Major Autohemotherapy (MAH) in cases of hyperglycemia. DDBD is a double cylinder made of borosilicate glass with a thickness of 0.2 cm and electrodes made of aluminum mesh with a length of 6.2 cm, an inner diameter of 2 cm and an outer diameter of 4.2 cm. The input gas is oxygen (O2) with flowrates of 0.2 L/minute-0.8 L/minute. At the flowrate limit used, this study found that the ozone concentration increased with increasing flowrate and operating voltage. The standard dose of MAH which is in accordance with the provisions of the Madrid Declaration for cases of diabetes mellitus/hyperglycemia, the specific dose is at the flow rate. of 0.4 L/min and a voltage of 3400 V, concentration of 293.64 μg/ml, capacity of 117.45 μg/second, with a dose obtained for the low dose of 18 μg and the high dose of 33 μg.
Keywords: Medical ozone, double dielectric barrier discharge, major autohemoteraphy (MAH), hiperglikemia.